@oimlsmart/primmel-packages 0.19.0 → 0.21.0
This diff represents the content of publicly available package versions that have been released to one of the supported registries. The information contained in this diff is provided for informational purposes only and reflects changes between package versions as they appear in their respective public registries.
- package/oiml-r129/execution/test-report-skeleton.prl +236 -0
- package/oiml-r144/execution/test-report-skeleton.prl +254 -0
- package/oiml-r60/execution/test-report-skeleton.prl +399 -0
- package/oiml-r91/execution/test-report-skeleton.prl +426 -0
- package/package.json +1 -1
- package/oiml-r129/payload/sample-data.yaml +0 -300
- package/oiml-r144/payload/sample-data.yaml +0 -643
- package/oiml-r60/payload/sample-data.yaml +0 -4900
- package/oiml-r91/payload/sample-data.yaml +0 -464
|
@@ -0,0 +1,399 @@
|
|
|
1
|
+
# OIML R 60 — test report skeleton (kernel 1.14.0's test_report_skeleton,
|
|
2
|
+
# TODO.roadmap/40): the R 60-3 type evaluation report structure — sections
|
|
3
|
+
# and the form inventory (C140 resolves the entries against the forms
|
|
4
|
+
# register; required/applicability carry the inclusion rules).
|
|
5
|
+
test_report_skeleton r60-3/test-report {
|
|
6
|
+
title "OIML R 60 Type Evaluation Report"
|
|
7
|
+
description "The complete evaluation report for an OIML R 60 load cell type evaluation, structured according to R 60-3. Contains administrative information, examination results, and performance test data."
|
|
8
|
+
section 4 {
|
|
9
|
+
title "Evaluation Report"
|
|
10
|
+
description "Administrative and descriptive information about the type evaluation"
|
|
11
|
+
form r60-3/sec-4.1 {
|
|
12
|
+
file "04-01-authority-info"
|
|
13
|
+
title "Issuing Authority"
|
|
14
|
+
required always
|
|
15
|
+
}
|
|
16
|
+
form r60-3/sec-4.2 {
|
|
17
|
+
file "04-02-synopsis"
|
|
18
|
+
title "Synopsis of Results"
|
|
19
|
+
required always
|
|
20
|
+
}
|
|
21
|
+
form r60-3/sec-4.3.1 {
|
|
22
|
+
file "04-03-01-examinations"
|
|
23
|
+
title "Summary of Examination Results"
|
|
24
|
+
required always
|
|
25
|
+
}
|
|
26
|
+
form r60-3/sec-4.3.2-a {
|
|
27
|
+
file "04-03-02-a-performance-general"
|
|
28
|
+
title "Summary of Performance Test Results (General)"
|
|
29
|
+
required always
|
|
30
|
+
}
|
|
31
|
+
form r60-3/sec-4.3.2-b {
|
|
32
|
+
file "04-03-02-b-performance-digital"
|
|
33
|
+
title "Summary of Performance Test Results (Digital)"
|
|
34
|
+
required conditional
|
|
35
|
+
applicability {
|
|
36
|
+
technology: [digital, digital-with-processing]
|
|
37
|
+
}
|
|
38
|
+
}
|
|
39
|
+
form r60-3/sec-4.4.1 {
|
|
40
|
+
file "04-04-01-manufacturer"
|
|
41
|
+
title "Manufacturer of the Specimen"
|
|
42
|
+
required always
|
|
43
|
+
}
|
|
44
|
+
form r60-3/sec-4.4.2 {
|
|
45
|
+
file "04-04-02-applicant"
|
|
46
|
+
title "Applicant Information"
|
|
47
|
+
required always
|
|
48
|
+
}
|
|
49
|
+
form r60-3/sec-4.4.3 {
|
|
50
|
+
file "04-04-03-testing-laboratories"
|
|
51
|
+
title "Testing Laboratories"
|
|
52
|
+
required always
|
|
53
|
+
}
|
|
54
|
+
form r60-3/sec-4.5 {
|
|
55
|
+
file "04-05-load-cell-type-info"
|
|
56
|
+
title "Load Cell Type Information"
|
|
57
|
+
required always
|
|
58
|
+
}
|
|
59
|
+
form r60-3/sec-4.6 {
|
|
60
|
+
file "04-06-accessories"
|
|
61
|
+
title "Accessories"
|
|
62
|
+
required always
|
|
63
|
+
}
|
|
64
|
+
form r60-3/sec-4.7 {
|
|
65
|
+
file "04-07-sample-selection"
|
|
66
|
+
title "Selection of Samples and Adjustments (§4.7–4.8)"
|
|
67
|
+
required always
|
|
68
|
+
}
|
|
69
|
+
form r60-3/sec-4.9.1 {
|
|
70
|
+
file "04-09-01-general-info"
|
|
71
|
+
title "Additional Information — General"
|
|
72
|
+
required always
|
|
73
|
+
}
|
|
74
|
+
form r60-3/sec-4.9.2 {
|
|
75
|
+
file "04-09-02-performance-info"
|
|
76
|
+
title "Additional Information — Performance Tests"
|
|
77
|
+
required always
|
|
78
|
+
}
|
|
79
|
+
form r60-3/sec-4.9.3 {
|
|
80
|
+
file "04-09-03-digital-info"
|
|
81
|
+
title "Additional Information — Digital Load Cells"
|
|
82
|
+
required conditional
|
|
83
|
+
applicability {
|
|
84
|
+
technology: [digital, digital-with-processing]
|
|
85
|
+
}
|
|
86
|
+
}
|
|
87
|
+
form r60-3/sec-4.9.4 {
|
|
88
|
+
file "04-09-04-photographs"
|
|
89
|
+
title "Photographs of Examinations and Tests"
|
|
90
|
+
required always
|
|
91
|
+
}
|
|
92
|
+
form r60-3/sec-4.9.5 {
|
|
93
|
+
file "04-09-05-documentation"
|
|
94
|
+
title "Documentation Supplied with the Test Pattern"
|
|
95
|
+
required always
|
|
96
|
+
}
|
|
97
|
+
form r60-3/sec-4.9.6 {
|
|
98
|
+
file "04-09-06-inscriptions"
|
|
99
|
+
title "Inscriptions and Presentations"
|
|
100
|
+
required always
|
|
101
|
+
}
|
|
102
|
+
form r60-3/sec-4.9.7 {
|
|
103
|
+
file "04-09-07-model-range"
|
|
104
|
+
title "Various Designs within the Model Range"
|
|
105
|
+
required always
|
|
106
|
+
}
|
|
107
|
+
form r60-3/sec-4.9.12 {
|
|
108
|
+
file "04-09-12-previous-tests"
|
|
109
|
+
title "Results of Previous Tests Taken into Account"
|
|
110
|
+
required always
|
|
111
|
+
}
|
|
112
|
+
form r60-3/sec-4.10.1 {
|
|
113
|
+
file "04-10-01-force-generating-system"
|
|
114
|
+
title "Test Equipment — Force Generating System"
|
|
115
|
+
required always
|
|
116
|
+
}
|
|
117
|
+
form r60-3/sec-4.10.2 {
|
|
118
|
+
file "04-10-02-weights"
|
|
119
|
+
title "Test Equipment — Weights"
|
|
120
|
+
required always
|
|
121
|
+
}
|
|
122
|
+
form r60-3/sec-4.10.3 {
|
|
123
|
+
file "04-10-03-temperature-chamber"
|
|
124
|
+
title "Test Equipment — Temperature Chamber"
|
|
125
|
+
required always
|
|
126
|
+
}
|
|
127
|
+
form r60-3/sec-4.10.4 {
|
|
128
|
+
file "04-10-04-climate-chamber"
|
|
129
|
+
title "Test Equipment — Climate Chamber"
|
|
130
|
+
required conditional
|
|
131
|
+
applicability {
|
|
132
|
+
humidity_class: [CH, SH]
|
|
133
|
+
}
|
|
134
|
+
}
|
|
135
|
+
form r60-3/sec-4.10.5 {
|
|
136
|
+
file "04-10-05-indicator"
|
|
137
|
+
title "Test Equipment — Indicator"
|
|
138
|
+
required always
|
|
139
|
+
}
|
|
140
|
+
form r60-3/sec-4.10.6 {
|
|
141
|
+
file "04-10-06-terminal"
|
|
142
|
+
title "Test Equipment — Terminal"
|
|
143
|
+
required conditional
|
|
144
|
+
applicability {
|
|
145
|
+
technology: [analogue-active, digital, digital-with-processing]
|
|
146
|
+
}
|
|
147
|
+
}
|
|
148
|
+
form r60-3/sec-4.10.7 {
|
|
149
|
+
file "04-10-07-barometric-pressure-meter"
|
|
150
|
+
title "Test Equipment — Barometric Pressure Meter"
|
|
151
|
+
required always
|
|
152
|
+
}
|
|
153
|
+
form r60-3/sec-4.10.8 {
|
|
154
|
+
file "04-10-08-thermometer"
|
|
155
|
+
title "Test Equipment — Thermometer"
|
|
156
|
+
required always
|
|
157
|
+
}
|
|
158
|
+
form r60-3/sec-4.10.9 {
|
|
159
|
+
file "04-10-09-moisture-analyser"
|
|
160
|
+
title "Test Equipment — Moisture Analyser"
|
|
161
|
+
required conditional
|
|
162
|
+
applicability {
|
|
163
|
+
humidity_class: [CH, SH]
|
|
164
|
+
}
|
|
165
|
+
}
|
|
166
|
+
form r60-3/sec-4.10.10 {
|
|
167
|
+
file "04-10-10-additional-test-equipment"
|
|
168
|
+
title "Test Equipment — Additional Equipment"
|
|
169
|
+
required always
|
|
170
|
+
}
|
|
171
|
+
form r60-3/sec-4.10.11 {
|
|
172
|
+
file "04-10-11-remarks"
|
|
173
|
+
title "Test Equipment — Remarks"
|
|
174
|
+
required always
|
|
175
|
+
}
|
|
176
|
+
form r60-3/form-1 {
|
|
177
|
+
file "test-usage-matrix"
|
|
178
|
+
title "Test Equipment Usage Matrix"
|
|
179
|
+
required always
|
|
180
|
+
}
|
|
181
|
+
}
|
|
182
|
+
section 5 {
|
|
183
|
+
title "Examinations"
|
|
184
|
+
description "Visual and documentation examinations per R 60-2 sections 2.5–2.6"
|
|
185
|
+
form r60-3/sec-5.1 {
|
|
186
|
+
file "marking-examination"
|
|
187
|
+
title "Marking Examination"
|
|
188
|
+
conformance_test /conf/examinations/inscriptions
|
|
189
|
+
requirements { /req/technical/mandatory-markings /req/technical/class-designation /req/technical/load-type-designation /req/technical/temperature-designation /req/technical/humidity-designation /req/technical/nlc-designation }
|
|
190
|
+
required always
|
|
191
|
+
}
|
|
192
|
+
form r60-3/sec-5.2 {
|
|
193
|
+
file "suitability-examination"
|
|
194
|
+
title "Suitability for Testing"
|
|
195
|
+
conformance_test /conf/examinations/suitability
|
|
196
|
+
requirements { /req/technical/mandatory-additional-info }
|
|
197
|
+
required always
|
|
198
|
+
}
|
|
199
|
+
form r60-3/sec-5.3 {
|
|
200
|
+
file "software-examination"
|
|
201
|
+
title "Software Examination"
|
|
202
|
+
conformance_test /conf/examinations/software
|
|
203
|
+
requirements { /req/technical/software }
|
|
204
|
+
required conditional
|
|
205
|
+
applicability {
|
|
206
|
+
technology: [analogue-active, digital]
|
|
207
|
+
}
|
|
208
|
+
}
|
|
209
|
+
form r60-3/sec-5.4 {
|
|
210
|
+
file "documentation-examination"
|
|
211
|
+
title "Documentation Examination"
|
|
212
|
+
conformance_test /conf/examinations/documentation
|
|
213
|
+
requirements { /req/technical/mandatory-additional-info }
|
|
214
|
+
required always
|
|
215
|
+
}
|
|
216
|
+
}
|
|
217
|
+
section 6 {
|
|
218
|
+
title "Performance Tests"
|
|
219
|
+
description "Metrological and environmental performance tests per R 60-2 sections 2.10.1–2.10.7"
|
|
220
|
+
form r60-3/sec-6.1 {
|
|
221
|
+
file "performance-test-summary"
|
|
222
|
+
title "Results of Performance Tests"
|
|
223
|
+
required always
|
|
224
|
+
}
|
|
225
|
+
form r60-3/sec-6.2 {
|
|
226
|
+
file "initial-tests"
|
|
227
|
+
title "Initial Tests and General Notes"
|
|
228
|
+
required always
|
|
229
|
+
}
|
|
230
|
+
subsection "Base Metrological Tests" {
|
|
231
|
+
description "Required for all load cells"
|
|
232
|
+
form r60-3/table-6.3-6.4 {
|
|
233
|
+
file "load-test"
|
|
234
|
+
title "Load Test Data (parameterized: 3 runs for C/D, 5 runs for A/B)"
|
|
235
|
+
conformance_test /conf/metrological-tests/measurement-error-repeatability-mdlo
|
|
236
|
+
requirements { /req/metrological/mpe /req/metrological/repeatability }
|
|
237
|
+
required conditional
|
|
238
|
+
applicability {
|
|
239
|
+
accuracy_class: [A, B, C, D]
|
|
240
|
+
}
|
|
241
|
+
}
|
|
242
|
+
form r60-3/table-6.5 {
|
|
243
|
+
file "load-cell-errors"
|
|
244
|
+
title "Load Cell Errors (E_L)"
|
|
245
|
+
conformance_test /conf/metrological-tests/measurement-error-repeatability-mdlo
|
|
246
|
+
requirements { /req/metrological/mpe }
|
|
247
|
+
required always
|
|
248
|
+
}
|
|
249
|
+
form r60-3/table-6.6 {
|
|
250
|
+
file "repeatability"
|
|
251
|
+
title "Repeatability Errors (E_R)"
|
|
252
|
+
conformance_test /conf/metrological-tests/measurement-error-repeatability-mdlo
|
|
253
|
+
requirements { /req/metrological/repeatability }
|
|
254
|
+
required always
|
|
255
|
+
}
|
|
256
|
+
form r60-3/table-6.7 {
|
|
257
|
+
file "temperature-mdlo"
|
|
258
|
+
title "Temperature Effect on MDLO (C_M)"
|
|
259
|
+
conformance_test /conf/metrological-tests/measurement-error-repeatability-mdlo
|
|
260
|
+
requirements { /req/metrological/temperature-effect-mdlo }
|
|
261
|
+
required always
|
|
262
|
+
}
|
|
263
|
+
form r60-3/table-6.8 {
|
|
264
|
+
file "creep-dr"
|
|
265
|
+
title "Creep (C_C) and DR (C_DR)"
|
|
266
|
+
conformance_test /conf/metrological-tests/creep
|
|
267
|
+
requirements { /req/metrological/creep /req/metrological/creep-20-30 /req/metrological/dr }
|
|
268
|
+
required always
|
|
269
|
+
}
|
|
270
|
+
form r60-3/table-6.9 {
|
|
271
|
+
file "barometric-pressure"
|
|
272
|
+
title "Barometric Pressure Effects (C_P)"
|
|
273
|
+
conformance_test /conf/metrological-tests/barometric-pressure
|
|
274
|
+
requirements { /req/metrological/barometric-pressure }
|
|
275
|
+
required conditional
|
|
276
|
+
notes "May not be necessary depending on the design of the load cell (R 60-3, 2.1.6 footnote) — required unless the manufacturer demonstrates the cell is pressure-insensitive"
|
|
277
|
+
}
|
|
278
|
+
}
|
|
279
|
+
subsection "Humidity Tests" {
|
|
280
|
+
description "Conditional on humidity class designation"
|
|
281
|
+
form r60-3/table-6.10.1 {
|
|
282
|
+
file "humidity-ch"
|
|
283
|
+
title "Humidity Effects CH (Cyclic)"
|
|
284
|
+
conformance_test /conf/metrological-tests/humidity-ch
|
|
285
|
+
requirements { /req/metrological/humidity-ch /req/metrological/mpe }
|
|
286
|
+
required conditional
|
|
287
|
+
applicability {
|
|
288
|
+
humidity_class: [CH]
|
|
289
|
+
}
|
|
290
|
+
}
|
|
291
|
+
form r60-3/table-6.10.2 {
|
|
292
|
+
file "humidity-sh"
|
|
293
|
+
title "Humidity Effects SH (Steady-State)"
|
|
294
|
+
conformance_test /conf/metrological-tests/humidity-sh
|
|
295
|
+
requirements { /req/metrological/humidity-sh }
|
|
296
|
+
required conditional
|
|
297
|
+
applicability {
|
|
298
|
+
humidity_class: [SH]
|
|
299
|
+
}
|
|
300
|
+
}
|
|
301
|
+
}
|
|
302
|
+
subsection "Electronic Load Cell Tests" {
|
|
303
|
+
description "Required for analogue-active and digital load cells only"
|
|
304
|
+
applicability {
|
|
305
|
+
technology: [analogue-active, digital]
|
|
306
|
+
}
|
|
307
|
+
form r60-3/form-6.11 {
|
|
308
|
+
file "warm-up-time"
|
|
309
|
+
title "Warm-Up Time"
|
|
310
|
+
conformance_test /conf/electronic-tests/warm-up-time
|
|
311
|
+
requirements { /req/electronic/warm-up-time }
|
|
312
|
+
required conditional
|
|
313
|
+
applicability {
|
|
314
|
+
technology: [analogue-active, digital]
|
|
315
|
+
}
|
|
316
|
+
}
|
|
317
|
+
form r60-3/form-6.12 {
|
|
318
|
+
file "power-voltage"
|
|
319
|
+
title "Power Voltage Variations"
|
|
320
|
+
conformance_test /conf/electronic-tests/power-voltage
|
|
321
|
+
requirements { /req/electronic/power-voltage-ac /req/electronic/power-voltage-dc /req/electronic/power-voltage-variations }
|
|
322
|
+
required conditional
|
|
323
|
+
applicability {
|
|
324
|
+
technology: [analogue-active, digital]
|
|
325
|
+
}
|
|
326
|
+
}
|
|
327
|
+
form r60-3/form-6.13 {
|
|
328
|
+
file "short-time-power"
|
|
329
|
+
title "Short-Time Power Reductions"
|
|
330
|
+
conformance_test /conf/electronic-tests/short-time-power
|
|
331
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
332
|
+
required conditional
|
|
333
|
+
applicability {
|
|
334
|
+
technology: [analogue-active, digital]
|
|
335
|
+
}
|
|
336
|
+
}
|
|
337
|
+
form r60-3/form-6.14 {
|
|
338
|
+
file "bursts"
|
|
339
|
+
title "Bursts (EFT)"
|
|
340
|
+
conformance_test /conf/electronic-tests/bursts
|
|
341
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
342
|
+
required conditional
|
|
343
|
+
applicability {
|
|
344
|
+
technology: [analogue-active, digital]
|
|
345
|
+
}
|
|
346
|
+
}
|
|
347
|
+
form r60-3/form-6.15 {
|
|
348
|
+
file "surge"
|
|
349
|
+
title "Surge"
|
|
350
|
+
conformance_test /conf/electronic-tests/surge
|
|
351
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
352
|
+
required conditional
|
|
353
|
+
applicability {
|
|
354
|
+
technology: [analogue-active, digital]
|
|
355
|
+
}
|
|
356
|
+
}
|
|
357
|
+
form r60-3/form-6.16 {
|
|
358
|
+
file "esd"
|
|
359
|
+
title "Electrostatic Discharge"
|
|
360
|
+
conformance_test /conf/electronic-tests/esd
|
|
361
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
362
|
+
required conditional
|
|
363
|
+
applicability {
|
|
364
|
+
technology: [analogue-active, digital]
|
|
365
|
+
}
|
|
366
|
+
}
|
|
367
|
+
form r60-3/form-6.17 {
|
|
368
|
+
file "emc-susceptibility"
|
|
369
|
+
title "Electromagnetic Susceptibility"
|
|
370
|
+
conformance_test /conf/electronic-tests/emc-susceptibility
|
|
371
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
372
|
+
required conditional
|
|
373
|
+
applicability {
|
|
374
|
+
technology: [analogue-active, digital]
|
|
375
|
+
}
|
|
376
|
+
}
|
|
377
|
+
form r60-3/form-6.18 {
|
|
378
|
+
file "conducted-emc"
|
|
379
|
+
title "Conducted Electromagnetic Fields"
|
|
380
|
+
conformance_test /conf/electronic-tests/conducted-emc
|
|
381
|
+
requirements { /req/electronic/disturbances /req/electronic/no-significant-faults }
|
|
382
|
+
required conditional
|
|
383
|
+
applicability {
|
|
384
|
+
technology: [analogue-active, digital]
|
|
385
|
+
}
|
|
386
|
+
}
|
|
387
|
+
form r60-3/form-6.19 {
|
|
388
|
+
file "span-stability"
|
|
389
|
+
title "Span Stability"
|
|
390
|
+
conformance_test /conf/electronic-tests/span-stability
|
|
391
|
+
requirements { /req/electronic/span-stability }
|
|
392
|
+
required conditional
|
|
393
|
+
applicability {
|
|
394
|
+
accuracy_class: [B, C, D]
|
|
395
|
+
}
|
|
396
|
+
}
|
|
397
|
+
}
|
|
398
|
+
}
|
|
399
|
+
}
|